Draeger, Inc. - DRA 6406550 - Chip Ammonia 0.2-5 ppm
Items #:
DRA 6406550
CMS-Chemical Specific Chip , For Dräger Chip Measurement System (Dräger CMS), Chemical – Ammonia, Measuring Range - 0.2-5 ppm, By Dräger, P/N 6406550
Draeger, Inc. - DRA 6718901 - DT Chloroprene 5/a (10)
Items #:
DRA 6718901
Product Specific Information; Dräger-Tubes - Chloroprene 5/a, Short-term Tubes, 10 tests per box, By Dräger, P/N 6718901
Standard Measuring Range : 5 to 60 ppm
Number of Strokes (n) : 3 + 3 desorptionstrokes in clean air.
Time for Measurement : app. 3 min
Standard Deviation : ± 10 to 15 %
Colour Change : violet -> yellow brown
Ambient Operating Conditions - Temperature : 10 to 30 °C,Absolute Humidity : 3 to 15 mg H2O / L
Reaction Principle - CH2=CH-CCI=CH2 + MnO4 – -> MnIV + various oxidation products
Cross Sensitivity - Many organic compounds with C=C double bonds are indicated,but with different sensitivities. It is not possible to differentiate between them.It is not possible to measure chloroprene in the presence of dialkylsulphides.
Draeger, Inc. - DRA 6726801 - DET.TUBE F. ETHYL GLYCOL ACETA
Items #:
DRA 6726801
Product Specific Information; Dräger Tubes - Ethyl Glycol Acetate 50/a, 50 - 700 ppm, Short-term Tubes, 10 tests per box, By Draeger, P/N 6726801
Standard Measuring Range : 50 to 700 ppm
Number of Strokes (n) : 10
Time for Measurement : app. 3 min
Standard Deviation : ± 20 to 30 %
Colour Change : yellow -> turquoise green
Ambient Operating Conditions - Temperature : 19 to 21 °C, For temperatures outside the 19 to 21oC range, multiply the tube reading by the following factor (K): T [°C] 10 15 18 22 25 30 35 K 2.1 1.8 1.3 0.9 0.7 0.6 0.5
Absolute Humidity : 5 to 12 mg H2O / L
Reaction Principle - Ethyl glycol acetate + CrVI -> CrIII + various oxidation products
Cross Sensitivity - Alcohols, esters, aromatics and ethers are also indicated, but withdifferent sensitivities. It is impossible to differentiate between them.
Draeger, Inc. - DRA 6727101 - DT Sulfur Dioxide 0.1/a (10)
Items #:
DRA 6727101
Product Specific Information; Dräger Tubes - Sulphur Dioxide 0.1/a, Short Term Tubes, 10 tests per box, Draeger P/N 6727101
Standard Measuring Range : 0.1 to 3 ppm
Number of Strokes (n) : 100
Time for Measurement : app. 20 min
Standard Deviation : ± 10 to 15 %
Colour Change : yellow -> orange
Ambient Operating Conditions
Temperature : 10 to 30 °C
Absolute Humidity : 3 to 15 mg H2O / L
Reaction Principle
SO2 + Na2[HgCI4] + Methyl red -> Na2[Hg(SO3)2] + 4 HCI
Cross Sensitivity