Return to previous page
Draeger, Inc. - DRA CH28101 - DT Hydrogen Sulfide 0,2%/A (10)
Item #: DRA CH28101
Product Specific Information; Dräger-Tubes - Hydrogen Sulphide 0.2%/A, Short-term Tubes, 10 tests per box, By Dräger, P/N CH28101
Standard Measuring Range : 0.2 to 7 Vol.-%
Number of Strokes (n) : 1 + 2 desorption strokesin air
Time for Measurement : app. 2 min
Standard Deviation : ± 5 to 10 %
Colour Change : pale blue -> black
Ambient Operating Conditions - Temperature : 0 to 60 °C
Absolute Humidity : max. 40 mg H2O / L
Reaction Principle - H2S + Cu2+ -> CuS + 2 H+
Cross Sensitivity - In the presence of sulphur dioxide, the indicating layer can change to a yellowish colour, but the hydrogen sulpide measurement is not affected. Comparable concentrations of mercaptan will interfere with the reading.
UNSPSC Code:
46160000
NIGP:
680.00000