Browse Categories
Return to previous page

Draeger, Inc. - DRA 8101341 - DR Tetrahydrothiophene 1/b (5)

Item #: DRA 8101341
Product Specific Information; Dräger Tubes - Tetrahydrothiophene 1/b, Short Term Tubes, 5 tests per box, Draeger P/N 8101341 Standard Measuring Range : 1 to 10 ppm Number of Strokes (n) : 30 Time for Measurement : app. 15 min app. 10 min for ameasurement in natural gas. Standard Deviation : ± 15 to 20 % Colour Change : violet -> yellow brown Ambient Operating Conditions Temperature : 0 to 35 °C Absolute Humidity : < 30 mg H2O / L Reaction Principle a) Adsorption with H2S b) THT + KMnO4 -> yellow brown reaction product Cross Sensitivity Up to 10 ppm hydrogen sulphide is adsorbed in the pretube,causing a brown discoloration. It is impossible to measure tetrahydrothiophene in the presence ofmercaptans. Up to 100 ppm of olefines (e.g. ethene, propene) will cause thecolour of the indicating layer to become lighter, at higher concentrationsthe olefins cause plus errors. Up to 200 ppm methanol does not interfere.

UNSPSC Code: 46160000
NIGP: 680.00000
$105.00

Call For Availability UOM : EA Qty:
Add to Quote