Browse Categories
Return to previous page

Draeger, Inc. - DRA 6733171 - DT Aniline 0.5/a (10)

Item #: DRA 6733171
Dräger-Tubes - Aniline 0.5/a, Short-term Tubes, 10 tests per box, By Dräger, P/N 6733171 Dräger-Tubes - Aniline 0.5/a, Short-term Tubes, 10 tests per box, By Dräger, P/N 6733171 Application Range Standard Measuring Range: 0.5 to 10 ppm Number of Strokes n: 20 Time for Measurement: app. 4 min Standard Deviation: ± 15 to 20 % Colour Change: pale yellow -> pale green Ambient Operating Conditions - Temperature: 15 to 30 °C,Absolute Humidity: 7 to 12 mg H2O / L Reaction Principle - C6H5NH2 + CrVI -> CrIII + various oxidation products Cross Sensivity - When methyl aniline and aniline are present at the same time, it isimpossible to measure only the aniline concentration.Ethers, ketones, esters, aromatics and petrols are indicated as well butwith different sensitivities.High aniline concentrations will not be indicated.

UNSPSC Code: 46160000
NIGP: 680.00000
$162.00

Call For Availability UOM : BX Qty: